A Low-Cost and Customizable TEER Meter for the Measurements of Cellular Barrier Integrity

dc.contributor.authorJones, Curtis G.
dc.contributor.authorChen, Chengpeng
dc.date.accessioned2020-01-31T18:32:28Z
dc.date.available2020-01-31T18:32:28Z
dc.date.issued2019-12-20
dc.description.abstractA low-cost and translational TEER (trans-endothelial/epithelial electric resistance) meter was designed, fabricated, validated, and applied in this paper. TEER is a critical tool to quantitate the integrity of biological barriers. Commercially available TEER meters are expensive (thousands of dollars) with low customization capability. Using Arduino, an open-source hardware and program that are used to control electronics, we fabricated the TEER meter that costs ~$50 to purchase the parts and 2 hours to be constructed. Robust characterization and validation shows that the meter can accurately measure TEER values between 132 and 82,500 Ω·cm² with <3% errors, which covers the reported TEER ranges based on a literature study we conducted. The temporal resolution, the measurement duration, and the electrode configurations of meter are also customizable. We successfully applied the meter to measure TEERs of endothelial cell monolayers, finding that cells treated with histamine have lower TEER values compared to untreated cells (793.4 ± 190.5 Ω·cm² vs. 3027.5 ± 664.4 Ω∙cm² ; p < 0.001), which is consistent with literature results. We further validated the TEER measurement by showing that histamine increased the intercellular gap from 2.34 ± 0.12 µm to 5.49 ± 0.17 µm, causing leakier endothelial barriers and thus lower TEERs. In conclusion, we report for the first time a low-cost Arduino-based TEER meter capable of accurately measuring TEERs in the relevant range. We also include detailed tutorials in the supplementary information to promote the translation of the technology.en
dc.description.urihttps://chemrxiv.org/articles/A_Low-Cost_and_Customizable_TEER_Meter_for_the_Measurements_of_Cellular_Barrier_Integrity/11374260/1en
dc.format.extent31 pagesen
dc.genrejournal articles preprintsen
dc.identifierdoi:10.13016/m2u6ij-njvf
dc.identifier.citationJones, Curtis G.; Chen, Chengpeng; A Low-Cost and Customizable TEER Meter for the Measurements of Cellular Barrier Integrity; ChemRxiv. Preprint.; https://chemrxiv.org/articles/A_Low-Cost_and_Customizable_TEER_Meter_for_the_Measurements_of_Cellular_Barrier_Integrity/11374260/1en
dc.identifier.urihttps://doi.org/10.26434/chemrxiv.11374260.v1
dc.identifier.urihttp://hdl.handle.net/11603/17203
dc.language.isoenen
dc.relation.isAvailableAtThe University of Maryland, Baltimore County (UMBC)
dc.relation.ispartofUMBC Chemistry & Biochemistry Department Collection
dc.relation.ispartofUMBC Student Collection
dc.relation.ispartofUMBC Student Collection
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0)*
dc.rightsThis item is likely protected under Title 17 of the U.S. Copyright Law. Unless on a Creative Commons license, for uses protected by Copyright Law, contact the copyright holder or the author.
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.titleA Low-Cost and Customizable TEER Meter for the Measurements of Cellular Barrier Integrityen
dc.typeTexten

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
A_Low-Cost_and_Customizable_TEER_Meter_for_the_Measurements_of_Cellular_Barrier_Integrity_v1.pdf
Size:
2.07 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
2.56 KB
Format:
Item-specific license agreed upon to submission
Description: