System performance of a TDM test-bed with long flex harness towards the new X-IFU FPA-DM

Department

Program

Citation of Original Publication

Rights

This work was written as part of one of the author's official duties as an Employee of the United States Government and is therefore a work of the United States Government. In accordance with 17 U.S.C. 105, no copyright protection is available for such works under U.S. Law.
Public Domain Mark 1.0

Abstract

SRON (Netherlands Institute for Space Research) is developing the Focal Plane Assembly (FPA) for Athena X-IFU, whose Demonstration Model (DM) will use for the first time a time domain multiplexing (TDM)-based readout system for the on-board transition-edge sensors (TES). We report on the characterization activities on a TDM setup provided by NASA Goddard Space Flight Center (GSFC) and National Institute for Standards and Technology (NIST) and tested in SRON cryogenic test facilities. The goal of these activities is to study the impact of the longer harness, closer to X-IFU specs, in a different EMI environment and switching from a single-ended to a differential readout scheme. In this contribution we describe the advancement in the debugging of the system in the SRON cryostat, which led to the demonstration of the nominal spectral performance of 2.8 eV at 5.9~keV with 16-row multiplexing, as well as an outlook for the future endeavours for the TDM readout integration on X-IFU's FPA-DM at SRON.